Type of Product:Device
Medsafe Reference:23854
Brand Name:ThermoCool SmartTouch SF Unidirectional
Model:
Affected:Catalogue No's: D134701, D134702, D134703, D134801, D134802, D134803, D134804, D134805
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Biosense Webster Inc
Brand Name:Thermocool SmartTouch SF Bidirectional catheter
Model:
Affected:Catalogue No's: D134701, D134702, D134703, D134801, D134802, D134803, D134804, D134805
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Biosense Webster Inc
Issue:The manufacturer has observed a higher than expected reported rate of char formation for the THERMOCOOL SMARTTOUCH® SF Uni-Directional Navigation Catheter and the THERMOCOOL SMARTTOUCH® SF Bi-Directional Navigation Catheter (collectively, the "THERMOCOOL SMARTTOUCH® SF Catheters"). Although char formation by itself is not a patient adverse event, in rare circumstances it has the potential to contribute to patient adverse events.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/11/2018