Detail of Recall

Type of Product:Device
Medsafe Reference:23853
Brand Name:CombiDiagnost R90
Model:All CombiDiagnost systems with software version 1.0.0, 1.0.1 and 1.0.2
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:Due to a software bug, the mA values are not locked, but increase when the operator restarts pulsed fluoroscopy several times after activation of the lock-in function. As a result, the patient received an increased radiation dose
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/11/2018