Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23840
Brand Name:W.O.M. GmbH - IIa - 60738; Endoscopic insufflation tubing set
Model:
Affected:PneumoClear Heated High-Flow Tube Set
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:WOM World of Medicine GmbH
Issue:A complaint was received regarding a puncture of the sterile pouch packaging due to a kinked tube set which orientated the prongs of the connection toward the pouch packaging.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/11/2018
Hide menus