Detail of Recall

Type of Product:Device
Medsafe Reference:23816
Brand Name:Intuitive Surgical - Instrument Drapes
Model:
Affected:D172997,D173077,D173517,D173607
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Intuitive Surgical Inc
Issue:During internal inspection, it was found that the da Vinci Si drape pouches may have been distributed with wrinkles along the pouch seal that could result in a channel within the seal, this channel may result in a breach in the sterility of the product.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/11/2018