Detail of Recall

Type of Product:Device
Medsafe Reference:23753
Brand Name:LivaNova Heater Cooler
Model:Heater-Cooler 3T, 230V
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:LivaNova Deutschland GmbH
Issue:The manufacturer is advising all 3T Heater Cooler users that following the current version of the cleaning instructions within the Instruction for Use (IFU). Additionally, the manufacturer is providing all 3T Heater Cooler users with updated instructions to monitor the concentration of hydrogen peroxide in the water circuit to verify that sufficient concentration of hydrogen peroxide is present to limit microbial growth, and to adjust the concentration of hydrogen peroxide if it drops below 100 ppm.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/10/2018