Detail of Recall

Type of Product:Device
Medsafe Reference:23752
Brand Name:Access hsTnI Reagent
Model:
Affected:Ref B52699 All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The Cardiac Troponin T cross reactivity concentration of 253 ng/mL (253,000 pg/mL) previously listed in the Access hsTnI Instructions For Use (IFU) does not apply to samples with hsTnI concentrations less than 16 pg/mL (ng/L).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/10/2018