Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23742
Brand Name:Anaesthesia Work Stations
Model:All upgraded Aisys anesthesia devices (i.e. Aisys devices upgraded to software version 11) and Aisys CS2 anesthesia devices (GTIN: 840682102292) with PSVPro option and/or End Tidal Control option.
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware that : Upgraded Aisys anesthesia devices (i.e. Aisys devices upgraded to software version 11) and Aisys CS2 anesthesia devices with End Tidal Control option (has the inability to set fresh gas settings) and/or PSVPro option (non-functional back-up ventilation after Cycling Procedure).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/10/2018
Hide menus