Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23731 |
| Brand Name: | Anaesthesia workstation systems |
| Model: | Fabius Anaesthesia Machine and the Auxiliary Common Gas Outlet. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | The manufacturer issued a safety Alert in November 2017 which addressed that the external fresh-gas outlet / auxiliary common gas outlet on Anaesthesia Workstations had, in some cases not been used in accordance to the Instructions for Use. The manufacturer has created a separate device label as an additional reminder to check the setting of the fresh-gas path when starting a mechanical Ventilation or ManSpont, (i.e. when using the COSY rebreathing System). The label consists of a graphical and a textual part/description and it is designed to fit to the Ventilator control panel next to the keys for selecting the Ventilation modes. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/10/2018 |