Type of Product:Device
Medsafe Reference:23725
Brand Name:ARCHITECT Free T3 Reagent Kit
Model:
Affected:All ARCHITECT Free T3 Reagent Kit - All lot numbers
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Abbott Australasia Pty Ltd
Issue:Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assay may show falsely depressed results when tested on board with select ARCHITECT assays and on board with the ARCHITECT 25-OH Vitamin D assay.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/10/2018