Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23703 |
| Brand Name: | Plum Primary Set |
| Model: | |
| Affected: | Lifeshield™ Primary Plum™ Set 15 Micron Filter - Lot no. 865045H |
| Software version: | |
| Recalling Organisation: | ICU Medical Australia Pty Ltd (NZ) Level 7 The Bayleys Building 36 Brandon Street Wellington 6011 |
| Contact Information: | |
| Manufacturer: | ICU Medical Inc |
| Issue: | Plum infusion pumps perform a Cassette Test when a new Plum Set is loaded into the pump. The manufacturer received reports of the pump not recognizing the cassette resulting in a Cassette Test Failure (N251) alarm, of Plum Sets distributed in New Zealand from June 2018 through August 2018. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/10/2018 |