Detail of Recall

Type of Product:Device
Medsafe Reference:23687
Brand Name:Hamburg DXR
Model:Affects all DigitalDiagnost systems with system software versions 4.1.x or 4.2.x and stitching license
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:When stitching image acquisition by means of the SkyPlate detector, the system software is designed such that the expected interval between two consecutive images is not smaller than eight seconds. If the detector readout time is faster so that the next image could be read in a time interval shorter than eight seconds, the synchronization between the detector and the system software results in a system abort of the stitching run. If this occurs, the stitching run needs to be repeated from the beginning.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/10/2018