Detail of Recall

Type of Product:Device
Medsafe Reference:23686
Brand Name:Kurz - Middle Ear Instrument Set
Model:
Affected:Instrumentation set: Product: 8000155 Lot: B21705026 - Affected component: Product: 8000110 Lot: B21705034
Software version:
Recalling Organisation:Teleflex Medical New Zealand
OFFICE 011, 110 CARLTON GORE ROAD
Newmarket
Auckland 1023
Contact Information:027 476 4009
Manufacturer:Heinz Kurz GmbH Medizintechnik
Issue:In some of the instrument sets the recess for placing the distance plates is to narrow. This results in an angulated position of the distance plates inside the cutting block. Therefore, the harvested cartilage that is placed on top of these distance plates cannot be cut evenly in thickness and in some cases cutting the cartilage might not be possible at all as the cutting blade could be blocked by the angulated distance plate.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/10/2018