Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23678 |
| Brand Name: | Samsung DR GC80 |
| Model: | |
| Affected: | Serial number - S086M3FG20002D (GC80) |
| Software version: | |
| Recalling Organisation: | Samsung Electronics New Zealand 24 The Warehouse Way, Northcote Auckland 0627 |
| Contact Information: | 09 4777000 |
| Manufacturer: | Samsung Medison Co Ltd |
| Brand Name: | Samsung DR GC85A |
| Model: | |
| Affected: | Serial number - 50FBM3GG900001B (GC85), 51EFM3GJ900001V (GC85), 51EFM3GJ900002W (GC85), 51EFM3GJ900003M (GC85) |
| Software version: | |
| Recalling Organisation: | Samsung Electronics New Zealand 24 The Warehouse Way, Northcote Auckland 0627 |
| Contact Information: | 09 4777000 |
| Manufacturer: | Samsung Medison Co Ltd |
| Issue: | The manufacturer has identified that, when the table of the GC80 or GC85 product is moved down, there is a risk of entrapment and injury in between the table legs. The second potential issue is that when the detector is not rebooted for at least one month, there is a risk of images being lost. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/10/2018 |