Detail of Recall

Type of Product:Device
Medsafe Reference:23671
Brand Name:RAPIDPoint® 500 Blood Gas Analyzer
Model:
Affected:V2.2, V2.3 and V2.3.1
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has identified an issue with the optional Custom Demographics software feature. This optional feature enables you to define up to 10 free text fields. The issue impacts the Custom Demographics fields sent to the LIS and viewed/printed from the data recall screen. These fields may get mixed up with Custom Demographics from another patient.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/10/2018