Type of Product:Device
Medsafe Reference:23669
Brand Name:Biograph mCT Hoffman Estates
Model:Biograph mCT and Biograph Horizon Systems produced between 11/2/2009 and 24/10/2017
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer has identifiied that power connectors on some devices may have fasteners which are not adequately tightened. This condition could cause the system power connection plug to overheat and lead to potential fire. The manufacturer will inspect all systems through a service visit to ensure power connector terminal screws are torqued to the proper specifications.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/10/2018