Detail of Recall

Type of Product:Device
Medsafe Reference:23646
Brand Name:RaySearch_Radiation therapy treatment planning system
Model:
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:The manufacture has identified an issue found with robust optimization in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7. When using robust optimization functions the background dose is not taken into account although in some cases it is displayed as such in the user interface.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/9/2018