Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23646 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacture has identified an issue found with robust optimization in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7. When using robust optimization functions the background dose is not taken into account although in some cases it is displayed as such in the user interface. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/9/2018 |