Type of Product:Device
Medsafe Reference:23644
Brand Name:Oscor Pacing Wire / Lead
Model:DP-07015 - DP-06060 - DP-05687 - DP-04455 - DP-03424
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Oscor Incorporated
Issue:During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss of connectivity or breakage during movement of the cables causing interruption of the pacing system.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:1/10/2018