| Type of Product: | Device |
| Medsafe Reference: | 23644 |
| Brand Name: | Oscor Pacing Wire / Lead |
| Model: | DP-07015 - DP-06060 - DP-05687 - DP-04455 - DP-03424 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Oscor Incorporated |
| Issue: | During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss of connectivity or breakage during movement of the cables causing interruption of the pacing system. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 1/10/2018 |