Detail of Recall

Type of Product:Device
Medsafe Reference:23637
Brand Name:Telemetry Systems Rimkus
Model:All B20/B20i/B40/B40i, and B105/B125 monitors
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Rimkus Medizintechnik GmbH
Issue:When multiple Bx0 or B1x5 patient monitors are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected. Loss of overall monitoring for a prolonged time may lead to a delay in detection of permanent or irreversible impairment or life-threatening changes in the condition of the patient
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/9/2018