Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23637 |
| Brand Name: | Telemetry Systems Rimkus |
| Model: | All B20/B20i/B40/B40i, and B105/B125 monitors |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Rimkus Medizintechnik GmbH |
| Issue: | When multiple Bx0 or B1x5 patient monitors are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected. Loss of overall monitoring for a prolonged time may lead to a delay in detection of permanent or irreversible impairment or life-threatening changes in the condition of the patient |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/9/2018 |