Detail of Recall

Type of Product:Device
Medsafe Reference:23614
Brand Name:Patient Monitor, Multi-Parameter
Model:All CARESCAPE Monitor B450, CARESCAPE Monitor B650, and CARESCAPE Monitor B850
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Issue:When multiple CARESCAPE Monitor B450, B650, or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/9/2018