Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23594
Brand Name:Dimension Vista® Lipase (LIPL)
Model:
Affected:Dimension Vista® Lipase (LIPL) - Lot numbers 18128BA, 18172BA, 18186BB
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed lot-to-lot variability for QC and patient results when using the Dimension Vista LIPL. The lot-to-lot bias observed for patient samples using the Dimension Vista lots when compared with a control lot ranged from 9% to 34%. Not all laboratories will observe this issue.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/9/2018
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /