Detail of Recall

Type of Product:Device
Medsafe Reference:23593
Brand Name: Dimension® Lipase (LIPL)
Model:
Affected:Dimension® Lipase (LIPL) - lot numbers GD9002, GA9003, GB9032, GA9038
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed lot-to-lot variability for QC and patient results when using the Dimension® LIPL lots. The lot-to-lot bias observed for patient samples using the affected Dimension® lots when compared with a control lot could be as high as 29%. Not all laboratories will observe this issue.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/9/2018