Type of Product:Device
Medsafe Reference:23556
Brand Name:ARCHITECT ICT Module
Model:
Affected:Lot Number 180326.
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Abbott Laboratories
Issue:The manufacturer has received customer complaints regarding higher than expected serum or plasma Chloride results when using ARCHITECT ICT Module. Based on an analysis of QC data from existing customers, the results generated using this lot number may be up to 6.2% higher than expected. The customer QC data do not indicate an issue with urine Chloride results.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/9/2018