Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23556 |
| Brand Name: | ARCHITECT ICT Module |
| Model: | |
| Affected: | Lot Number 180326. |
| Software version: | |
| Recalling Organisation: | Abbott Laboratories (NZ) Ltd Unit D 4 Pacific Rise Mt Wellington Auckland 2025 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Laboratories |
| Issue: | The manufacturer has received customer complaints regarding higher than expected serum or plasma Chloride results when using ARCHITECT ICT Module. Based on an analysis of QC data from existing customers, the results generated using this lot number may be up to 6.2% higher than expected. The customer QC data do not indicate an issue with urine Chloride results. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/9/2018 |