Brand Name: Zebra PSU
Model:Zebra TLP3842 effected - part number 808113-001 or P1028888-001, Zebra GX430t - 808101-001
Affected:
Software version:
Recalling Organisation:Leica Microsystems Pty Ltd, New Zealand Branch
3/33 Spartan Road
Takanini
Auckland 2105
Contact Information:021 0841 7323
Manufacturer:Nano-X AI Ltd
Issue:The Zebra Printer PSUs for the affected models of Zebra printers detailed above, which serve as the external power source for these printers, have been identified as potentially containing an improperly mixed phosphorus compound that is used in the flame retardant resin in the DC power supply connectors. This deficiency in the phosphorus based connector resin, when combined with moisture and humidity over time, can potentially lead to overheating or a fire hazard.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/9/2018