Detail of Recall

Type of Product:Device
Medsafe Reference:23537
Brand Name:Haemodialysis system - Lundia AB
Model:AK98 Haemodialysis Devices
Affected:
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Gambro Lundia AB
Issue:The manufacturer has identified a potential for AK 98 Haemodialysis Devices to generate excessive ultrafiltration (UF) in certain situations where treatment-related alarms occur, or where there is an ultrafilter leak. Excessive UF may present hazards for sensitive patients, such as low-weight patients, for whom target UF values of zero or very low volumes are desired. Due to the potential for excess fluid loss, when treating low-weight or other sensitive patients, weight loss should be monitored during treatment and the attached mitigating instructions should be followed.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/8/2018