Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23535
Brand Name:CYPASS
Model:
Affected:All lots
Software version:
Recalling Organisation:Alcon Laboratories Australia Pty Ltd (NZ)
c/- Pharmaco (NZ) Ltd
4 Fisher Crescent
Mt Wellington
Auckland
Contact Information:09 270 4360
Manufacturer:Transcend Medical Inc
Issue:This Medical Device Recall is not related to a manufacturing or quality issue. Rather, this action has been initiated based on an analysis of the completed dataset from the COMPASS-XT long-term safety study. The analysis showed that the CyPass® Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:4/9/2018