Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23527 |
| Brand Name: | SM ASD Inc Southington - IV Catheters- 34920 |
| Model: | |
| Affected: | Jelco® ViaValve™ safety I.V. catheter Lots - 3584467, 3584468, 3584469, 3576073, 3576233, 3578511, 3578512 |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical ASD Inc |
| Issue: | The manufacturer identified that certain Jelco ViaValve safety I.V. catheters exhibited a V-shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process. Upon completion of the investigation, equipment readjustment was performed which immediately corrected the issue. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/8/2018 |