Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23527
Brand Name:SM ASD Inc Southington - IV Catheters- 34920
Model:
Affected:Jelco® ViaValve™ safety I.V. catheter Lots - 3584467, 3584468, 3584469, 3576073, 3576233, 3578511, 3578512
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical ASD Inc
Issue:The manufacturer identified that certain Jelco ViaValve safety I.V. catheters exhibited a V-shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process. Upon completion of the investigation, equipment readjustment was performed which immediately corrected the issue.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/8/2018
Hide menus