| Type of Product: | Device |
| Medsafe Reference: | 23506 |
| Brand Name: | Stryker GmbH - IIb - 35647 - Fixation device, external, system |
| Model: | |
| Affected: | Item Number - 49350010, 49350020, 49350030, 49350040, 49359900-T |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker GmbH |
| Issue: | The manufacturer has determined that products have left the factory where the press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/8/2018 |