Type of Product:Device
Medsafe Reference:23506
Brand Name:Stryker GmbH - IIb - 35647 - Fixation device, external, system
Model:
Affected:Item Number - 49350010, 49350020, 49350030, 49350040, 49359900-T
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker GmbH
Issue:The manufacturer has determined that products have left the factory where the press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory