Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23496 |
| Brand Name: | AFX & AFX2 EndoVascular AAA system |
| Model: | AFX Endovascular AAA System - All models |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Endologix New Zealand Co. Level 27 88 Shortland Street Auckland 1010 |
| Contact Information: | |
| Manufacturer: | Endologix International BV |
| Issue: | The manufacturer has conducted investigations into reports of Type III endoleaks which may cause increased pressure within the aneurysm sac that could increase the risk of aneurysm rupture and patient death. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/8/2018 |