| Type of Product: | Device |
| Medsafe Reference: | 23449 |
| Brand Name: | Arctic Sun |
| Model: | |
| Affected: | Lot numbers 5439 - DYASY023 - DYBNY110 - DYZJY044 |
| Software version: | |
| Recalling Organisation: | Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics 54 Carbine Road Mt Wellington AUCKLAND |
| Contact Information: | 09 969 0705 |
| Manufacturer: | Medivance Inc |
| Issue: | An improper wiring connection on the chiller assembly has been identified in the Bard Arctic Sun 5000 Temperature Management Systems which may result in overheating of the connection and potential failure of certain components in the system. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/8/2018 |