Type of Product:Device
Medsafe Reference:23443
Brand Name:Stryker GmbH - Fixation device, external, system
Model:
Affected:983031 Hoffmann LRF Hexapod Software HLRF 1.1-2/002 and 983031 Hoffmann LRF Hexapod Software HLRF 1.0-2/002
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker GmbH
Issue:The manufacturer has identified a deviation with the Hoffmann LRF Hexapod Software that occurs when both the Correct Axial First feature is selected and more than one correction per day is selected when computing a correction plan. The deviation results in the calculation of an accelerated correction plan for the axial portion of the plan.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/8/2018