Detail of Recall

Type of Product:Device
Medsafe Reference:23442
Brand Name:UniCel DxH 800 Coulter Cellular Analysis System
Model:3.2.0 and below, 1.3.0 and below
Affected:
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Australia Pty Ltd
Issue:This notification is an update to the previous urgent Product Correction Notice dated July 30, 2018 concerning sporadic erroneously elevated platelet results. The manufacturer has confirmed complaints of erroneously elevated platelet results with rare occurrences, without flags or system messages. The underlying issue is temporary disturbance of the sweep flow. Preliminary root cause investigation indicates that sweep flow disruption may occur following the Clear RBC Apertures procedure. The issue may affect one or multiple samples tested in sequence.
Recall action type:Recall
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/5/2019