Detail of Recall

Type of Product:Device
Medsafe Reference:23436
Brand Name:Fresenius 4008 Bloodlines
Model:4008 Bloodlines
Affected:
Software version:
Recalling Organisation:Fresenius Medical Care Australia Pty Ltd (NZ)
12D Rennie Drive
Mangere
AUCKLAND 2022
Contact Information:09 925 2700
Manufacturer:Fresenius Medical Care AG
Issue:A number of leakages were reported at several locations on the 4008 bloodline tubing systems. Leakages on the bloodlines may cause blood loss into the environment and may lead to air intake into the system. While the air intake is normally detected by the dialysis machine's air detector, a leakage of blood may not be detected.
Following the April 2018 Safety Alert (Medsafe Ref: 22357) issued for the 4008 bloodlines, the manufacturer has worked to change the 4008 bloodlines manufacturing process and has developed unaffected stock.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/8/2018