Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23431 |
| Brand Name: | MITG - Capnograph and Pulse Oximeter Monitor (Oridion) |
| Model: | Capnostream™ 20, Capnostream™ 20P |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Oridion Medical 1987 Ltd |
| Issue: | The manufacturer is issuing a software update for Capnostream™20 and Capnostream™20p Bedside Patient Monitors, following receipt of customer reports that the date/time, nurse call and alarm settings of these devices may be reset to the factory default settings when the monitor is powered off. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/8/2018 |