Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23425 |
| Brand Name: | Angiographic X-Ray |
| Model: | FlexVision Monitor Ceiling Suspension in the period 2003 to May 2011 are affected. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | The manufacturer received a complaint reporting that a Monitor Ceiling Suspension (MCS) with a FlexVision 56-Inch large screen fell to the ground. The actuator assembly of the MCS became detached and the monitor carriage with the FlexVision monitor dropped to the ground. The Monitor Ceiling Suspension is designed to allow flexible positioning near the patient table when in use, and away from the patient table when not in use (parked position). |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/8/2018 |