Detail of Recall

Type of Product:Device
Medsafe Reference:23425
Brand Name:Angiographic X-Ray
Model:FlexVision Monitor Ceiling Suspension in the period 2003 to May 2011 are affected.
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer received a complaint reporting that a Monitor Ceiling Suspension (MCS) with a FlexVision 56-Inch large screen fell to the ground.
The actuator assembly of the MCS became detached and the monitor carriage with the FlexVision monitor dropped to the ground.
The Monitor Ceiling Suspension is designed to allow flexible positioning near the patient table when in use, and away from the patient table when not in use (parked position).
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/8/2018