Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23411
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model: ARTIS Q and ARTIS Pheno
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:In the affected Artis system the temperature of the detector could leave its designed working range (cool down) due to a wrongly configured temperature control unit. As as result the affected Artis system may take longer time (up to 90 minutes) from system start up until it is possible to acquire clinical images.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/8/2018