Type of Product:Device
Medsafe Reference:23406
Brand Name:ARCHITECT EBV VCA IgM reagent
Model:
Affected: lots 85278LI00 and 85279LI00
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Diagnostics
Issue:The manufacturer has identified the two ARCHITECT EBV VCA IgM Reagent lots listed above may show reduced RLU (Relative Light Units) signal when used in combination with certain lots of ARCHITECT EBV VCA IgM Calibrator (LN 3P66-01). This reduced RLU signal may cause up to a 20% decrease in cutoff RLU values in assay calibration resulting in:
Upward shift in Quality Control S/CO values
False grayzone patient results in the range of 0.50-0.60 S/CO
False reactive patient results in the range 1.00-1.20 S/CO
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/7/2018