Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23399 |
| Brand Name: | Arrow International - Central Venous Catheter Sets |
| Model: | |
| Affected: | Lot # 14F15H0230 |
| Software version: | |
| Recalling Organisation: | Teleflex Medical New Zealand OFFICE 011, 110 CARLTON GORE ROAD Newmarket Auckland 1023 |
| Contact Information: | 027 476 4009 |
| Manufacturer: | Arrow International Inc |
| Issue: | The manufacturer has identified that the Arrow® HANDS-OFF Multi-Lumen Central Venous Catheter with Blue FlexTip packaging may not be sealed. If the packaging is compromised in this manner, the sterility of the product cannot be guaranteed. If a non-sterile product is used, there is potential for infection to occur. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/7/2018 |