Detail of Recall

Type of Product:Device
Medsafe Reference:23396
Brand Name:Steris - VPro sterilizer
Model:CABINET VPRO VHP UNIT - VPRO 1 PLUS SGL DOOR CABINET 400V 50HZ - VPRO 1 PLUS DD DOOR REC 1WALL 400V 50HZ - VPRO MAX SINGLE CABINET 400 50HZ ENGLISH - VPROMAX DBL REC 1WALL 400 50HZ ENGL
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Steris Corporation
Issue:The manufacturer has identified that in the unlikely scenario where a punctured cup of VAPROX HC Sterilant expires while the V-PRO unit (1, 1 Plus, or maX) is powered down, upon power-up the V-PRO unit will not detect the expired cup of sterilant; allowing for subsequent cycles to be run.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/7/2018