Detail of Recall

Type of Product:Device
Medsafe Reference:23394
Brand Name:MITG - Patient monitor, anaesthesia depth
Model:Li-ion battery packs used in BIS™ Vista Monitoring Systems
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Issue:The manufacturer has received reports of thermal damage due to the potential for aged battery packs to experience an internal short that may result in a thermal event. The manufacturer is issuing a correction to revise the replacement instructions for the Li-ion battery packs used in BIS™ Vista Monitoring Systems. These revised instructions will require battery pack replacement after 4 years.
Manufacturers actions:
New battery labels will include an expiration date
An update to the Operator's Manual will be initiated in phase two of this field action.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/8/2018