Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23394 |
| Brand Name: | MITG - Patient monitor, anaesthesia depth |
| Model: | Li-ion battery packs used in BIS™ Vista Monitoring Systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Covidien LLC |
| Issue: | The manufacturer has received reports of thermal damage due to the potential for aged battery packs to experience an internal short that may result in a thermal event. The manufacturer is issuing a correction to revise the replacement instructions for the Li-ion battery packs used in BIS™ Vista Monitoring Systems. These revised instructions will require battery pack replacement after 4 years. Manufacturers actions: New battery labels will include an expiration date An update to the Operator's Manual will be initiated in phase two of this field action. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/8/2018 |