Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23392 |
| Brand Name: | SureSigns |
| Model: | All SureSigns Monitors manufactured up to May 3, 2018 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | The manufacturer has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Update 30 January 2019: The manufacturer has released a system update and will contact customers with affected devices to facilitate installation of the software upgrade. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/1/2019 |