Detail of Recall

Type of Product:Device
Medsafe Reference:23392
Brand Name:SureSigns
Model:All SureSigns Monitors manufactured up to May 3, 2018
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles.
Update 30 January 2019: The manufacturer has released a system update and will contact customers with affected devices to facilitate installation of the software upgrade.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/1/2019