Detail of Recall

Type of Product:Device
Medsafe Reference:23384
Brand Name: RAPIDPoint Systems & RAPIDLab Systems
Model: RAPIDPoint 405 Blood Gas Analyzer - RAPIDPoint 500 Blood Gas Analyzer - RAPIDPoint 1245 Blood Gas Analyzer- RAPIDPoint 1265 Blood Gas Analyzer
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has determined that therapeutic levels (1 mg/mL and 2 mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the above-mentioned RAPIDPoint and RAPIDLab Blood Gas Systems.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/7/2018