Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23371 |
| Brand Name: | Intuitive - Endoscopic control system |
| Model: | Serial number - SK1230 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Device Technologies New Zealand 47 Arrenway Drive Albany AUCKLAND 0632 |
| Contact Information: | 09 913 2000 |
| Manufacturer: | Intuitive Surgical Inc |
| Issue: | The manufacturer has become aware of certain Redundant Medical Grade Power Supply (RMGPS) units in specific da Vinci Xi® and X® Surgeon Consoles that have been shipped with a manufacturing defect. The RMGPS is the primary power supply to the console and is designed with redundancy so that if one power supply module fails, the power load is shifted to the functioning power supply module. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/7/2018 |