Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23367 |
| Brand Name: | ARCHITECT Creatine Kinase (CK) regent |
| Model: | |
| Affected: | Lots 87697UN18, 99632UN18, 24358UN18 and 99633UN18 |
| Software version: | |
| Recalling Organisation: | Abbott Australasia Pty Ltd (NZ) Abbott Laboratories NZ 4 Pacific Rise, Mt Wellington Auckland 1060 |
| Contact Information: | 61 2 93849717 |
| Manufacturer: | Abbott Laboratories Limited |
| Issue: | The manufacturer has identified a stability issue that may lead to the following issues: · Error Code 1054 Unable to calculate result, reaction check failure · Quality Control (QC) results out of range low |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/7/2018 |