Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23352
Brand Name:IH-Com Fullversion
Model:
Affected:
Software version:version 5.1 only
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:The manufacturer has identified a software anomaly in IH-Com version 5.1.
This anomaly only affects the QC module for ABO forward grouping of IH-Com. The software allows the release of a QC result ("QC accepted") which is discrepant with the claimed result in the conditions described below:
1. Process ABO forward grouping tests with QC samples.
2. The QC result is not as expected due to a potential failure of instrument or reagents.
3. Obtain a discrepant result with the claimed result.
4. IH-Com displays a "QC passed" while it should be displayed as "QC failed".
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/7/2018