| Type of Product: | Device |
| Medsafe Reference: | 23302 |
| Brand Name: | ABL800 series analysers |
| Model: | All with software versions 6.14 through to Version 6.18 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | In July 2018, the manufacturer advised that the ABL800 series analysers could potentially provide a mix-up of patient results when Confirmation of Demographics is enabled as part of the query to AQURE, Radiance, or a HIS LIS system. (Note: The default setting is that this function is disabled.) The manufacturer has since been working on a permanent solution which involved a redesign of the analyser software. The permanent countermeasure to this issue is now released in software version 6.18 MR1. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/5/2019 |