Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23292 |
| Brand Name: | BACTALERT VIRTUO Instrument |
| Model: | BACT/ALERT® VIRTUO® A Unit, BACT/ALERT® VIRTUO® B Unit, BACT/ALERT® VIRTUO® A Unit China |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medi'Ray New Zealand Ltd 53-55 Paul Matthews Road Albany Auckland 0632 |
| Contact Information: | 0800 633 472 |
| Manufacturer: | Biomerieux SA |
| Issue: | The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/7/2018 |