Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23262 |
| Brand Name: | Hamburg DXR |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | The footplate of the Patient support for stitching can be fixed in upper position. When the fixing clamp (hook) is not in its correct locking position, the footplate is not securely held. If it falls down, it may hurt the operator's or patient's foot or leg (e.g. when moving the stand or when the operator fails to fix it correctly upon moving the footplate up). |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/6/2018 |