Detail of Recall

Type of Product:Device
Medsafe Reference:23262
Brand Name:Hamburg DXR
Model:
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The footplate of the Patient support for stitching can be fixed in upper position. When the fixing clamp (hook) is not in its correct locking position, the footplate is not securely held. If it falls down, it may hurt the operator's or patient's foot or leg (e.g. when moving the stand or when the operator fails to fix it correctly upon moving the footplate up).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/6/2018