Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23220 |
| Brand Name: | Artis Q, Q.zen, zeego, zee systems |
| Model: | |
| Affected: | Catalogue number – 10848282, 10848281, 10848280, 10848283, 10848355, 10849000 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | In case of a short-circuit within the X-ray tube a leakage current might cause an overload of the X-ray generator. If this leakage current keeps undetected it may damage the generator by overheating. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/6/2018 |