Detail of Recall

Type of Product:Device
Medsafe Reference:23177
Brand Name:Fujifilm ProSolv (Synapse CV) Cardio Software
Model:versions 4.0.2 through version 6.0.4.1 (Synapse CV).
Affected:
Software version:
Recalling Organisation:Fujifilm NZ Limited
2c William Pickering Drive
Albany
Auckland
Contact Information:0800 242 646
Manufacturer:Fujifilm Medical Systems USA Inc
Issue:The manufacturer has become aware of a malfunction in the system where inconsistent Mitral Valve Mean Pressure Gradient values are being reported. The values are only affected when measuring multiple velocity traces in patients with an irregular heart rhythm, with mitral and tricuspid valve stenosis, using the Synapse CV measurement tools. Multiple traces measured on the modality and transmitted to Synapse CV are not affected.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/6/2018