Detail of Recall

Type of Product:Device
Medsafe Reference:23145
Brand Name: MMC 2433 In-line blood set with filter
Model:
Affected:Multiple Lot numbers
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxalta Belgium Manufacturing SA
Issue:The manufacturer is issuing a recall for In-Line Blood Set with Filter, as they have the potential to leak at the tube to cap junction. Preliminary Investigation has shown that this issue is due to the combination of cap dimension variability, especially on lower limit, and insertion depth of the cap into the tube on the lower side.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2018